Operation and failuresGuide

Biogas analyser: calibration, drift of CH4, CO2, O2 and H2S sensors and verification of measurements

An analyser shows the composition of biogas, but its readings drift over time. We describe how the sensors work, calibration and verification, and the effect of CH₄ error.

Published 4 min readFeedstockData Editorial Team

Other language versions: Polski, Deutsch, Español, Français, Italiano, Nederlands.

Niebieski statyw z probówkami typu Eppendorf i pojemnik z końcówkami do pipet.
Photo Gannu03 / Wikimedia Commons, licence CC BY-SA 4.0, source file; changes: cropping

A biogas analyser is an instrument that the operator trusts uncritically, because it gives numbers to fractions of a percent. Meanwhile the sensors age, the sampling line clogs with condensate and calibration is put off. The result: an apparently stable reading that departs from reality by 1–2 percentage points. This guide concerns the interpretation and checking of composition measurement; the volume of gas is described in the article on flow meters and pressure compensation, and variation in composition results from the process and the substrates. H₂S appears here only as a measured parameter; its sources are described in the article on hydrogen sulphide in the digester.

How the sensors work

  • NDIR (non-dispersive infrared) measures the absorption of radiation by CH₄ and CO₂ at characteristic wavelengths. It is stable, but sensitive to fouling of the optics and condensation.
  • Electrochemical sensors (O₂, H₂S) generate a current proportional to the gas concentration. They wear out over time, their sensitivity falls, and the presence of other compounds can cause so-called sensor poisoning or cross-sensitivity.
  • Thermal conductivity sensors are sometimes used as a supplement; they respond to changes in gas composition, but do not distinguish components without additional information.

Accuracy, ranges and resistance depend on the model, so the values for a specific instrument are given in the manufacturer's data sheet.

Drift and poisoning

Drift is a slow change in the reading at a constant concentration. In NDIR it results mainly from fouling and ageing of the light source, in electrochemical sensors from depletion of the electrolyte and the action of contaminants. Poisoning is the result of exposure to concentrations above the range or to compounds that damage the active element; the reading falls, and sometimes the sensor stops responding. A characteristic symptom: the oxygen value "stands still" despite changes in the plant, or H₂S shows zero right after a change of conditions in the process.

Calibration

Calibration consists of comparing the readings with a gas of known concentration:

  1. Zero. A gas with zero content of the measured component (often nitrogen) sets the reference point.
  2. Reference gas (span). A mixture of known composition, close to the measured range (e.g. a CH₄/CO₂ mixture for NDIR).
  3. Record. The date, the value before and after calibration, the cylinder number and the expiry date of the gas.

The frequency is recommended by the manufacturer, usually at intervals of months; sensors working in harsh conditions may require more frequent checks. Reference gas has an expiry date and storage conditions, and after it is exceeded the composition may change. It is also worth checking that calibration is performed on the same line as normal measurement, to account for the effect of filters and tubing.

Sampling line

Many problems lie upstream of the instrument:

  • condensate in the sampling tube, which dissolves CO₂ and H₂S, changing the composition of the gas reaching the sensor,
  • clogged filters, restricting the sample flow,
  • leaks that let air in (O₂ rises, CH₄ falls in the measurement),
  • tubing that is too long and dead volumes, which delay the response,
  • no sample cooler or dryer, if the instrument requires one.

A typical leak test: a change in O₂ and N₂ with an unchanged process indicates air being drawn into the line.

Effect of the error on energy

The calorific value of biogas rises almost linearly with CH₄ concentration. For a gas with a concentration of about 55%, an overstatement of the reading by 1–2 percentage points means about 2–4% relative error in energy. For a plant with constant production this means a systematic error in the energy balance and in the assessment of CHP unit efficiency. The example is illustrative; the actual effect depends on the composition and the method of calculation.

Verification

The readings of a permanent analyser are worth periodically comparing with another source:

  • a portable analyser calibrated separately,
  • laboratory analysis of a sample taken in a bag or cylinder (gas chromatography),
  • balance consistency: the concentrations of CH₄ and CO₂ together with O₂ and N₂ should add up to a value close to 100%.

A discrepancy exceeding the uncertainty of the instrument (usually a fraction of a percentage point, and more for ageing sensors) is a signal to check the calibration, the sampling line and the condition of the sensors.

In daily work a simple rule is useful: every change in readings that has no explanation in the process is first suspected to be a measurement problem. For example, a jump in O₂ after a filter replacement is usually a leak, and a jump in CH₄ after calibration is the result of correcting earlier drift. Therefore after every calibration and replacement of a line element the readings before and after should be recorded. Such notes later make it possible to distinguish a true change in biogas composition from a correction of error, which matters in assessing the process and in energy settlements.

Key takeaways

An analyser requires the same attention as process equipment: regular calibration, line checks and independent verification. An error of 1–2 percentage points of CH₄ is a few percent of energy. Before interpreting a change in composition as a biological problem, it is worth making sure that the reading is reliable.